Process Trace Oxygen Analyzer

Model: 3020T

Brand: Teledyne / USA

Process Trace Oxygen Analyzer
  • An explosion proof NEMA 4/7 rated enclosure
  • A 2-line alphanumeric display screen, driven by microprocessor electronics continuously prompting and informing the operator
  • High resolution, accurate readings of oxygen content from low ppm levels through 25%
  • Advanced Micro-fuel Cell designed for trace analysis, with a one year warranty and an expected life of two years
  • Versatile analysis over a wide range of applications
  • Microprocessor based electronics: 8-bit CMOS microprocessor with 32kB RAM and 128kB ROM
  • Three user-definable output ranges (from 0-10ppm through 0 ~ 250,000 ppm) assuring a perfect match for the user's process
  • Air-calibration range for convenient spanning at 20.9%
  • Auto-Ranging automatically selects the proper preset range for a given measurement. Manual override allows the user to lock onto a specific range of interest
  • Two adjustable concentration alarms and a system failure alarm
  • Extensive self-diagnostic testing at start-up and on demand with continuous power-supply monitoring
  • Two way RS-232 serial digital port for use with a computer or other digital communication device
  • Four analog outputs; two for measurement (0-1 VDC and Isolated 4-20 mADC) and two for range identification